depends upon the diffusion coefficient in the liquid stationary phase.

      In practice, the values for the coefficients A, B, and C can be obtained by making several measurements of efficiency for the same compound undergoing chromatography at different flow rates, since flow rate and average linear speed are related. Next, the hyperbolic function that best satisfies the experimental values can be calculated, preferably by using the multiple linear regression method.

      1.10.2 Golay Equation

      In 1958, Golay proposed a modified relationship reserved for capillary columns used in gas chromatography:

      (1.40)upper H Subscript t h e o period min Baseline equals r StartRoot StartFraction 1 plus 6 k plus 11 k squared Over 3 left-parenthesis 1 plus k right-parenthesis squared EndFraction EndRoot

      

      1.10.3 Knox Equation

      (1.41)h equals upper A u overbar Superscript 1 slash 3 Baseline plus StartFraction upper B Over u overbar EndFraction plus upper C u overbar

      The utility of analytical chromatography depends on its ability to separate analytes, whether to identify or quantify them. Particularly in the latter case, peak areas must be defined with precision. Since many parameters affect the quality of results, there are a certain number of requirements regarding the resolution between solute peaks, the speed of analysis (elution time, then the time needed to return to the initial conditions of analysis if different from the final conditions), and even the column’s capacity (to avoid saturation). To do so, we often use software to simulate the separation by adjusting various physico‐chemical factors such as temperature, flow rate, and composition of the mobile phase.

Schematic illustration of the chromatographer’s compromise triangle between resolution, speed, and capacity.

      Often, in a chromatographic assay, we are only interested in one or two compounds present in the sample. If we have a selective detector adapted to the analytes in question, we can facilitate the optimization stage, since the resulting chromatogram will be greatly simplified. In other cases, however, we try to separate the greatest possible number of compounds from the initial mixture.

      Chromatographic techniques can be classified according to various criteria: the physical nature of the phases, the process used, or physico‐chemical phenomena occurring during separation, etc. The following classification has been established by consideration of the physical nature of the two phases involved.

      1.12.1 Liquid Chromatography (LC)

      This type, in which the mobile phase is a liquid, is the oldest known form of chromatography, with the earliest applications involving preparative methods of separation (for which it is still used today). Depending on the nature of the stationary phase (nature of interactions determining the separation of solutes), we can distinguish between liquid/solid chromatography (LSC), liquid/liquid chromatography (LLC), ion chromatography (IC), and size exclusion chromatography (SEC) (also known as gel permeation or gel filtration chromatography depending on the nature of the stationary phase).

      Liquid/solid chromatography (LSC)

      The stationary phase is a permeable solid medium to which the solute molecules adhere through the dual effect of physisorption and chemisorption. The physico‐chemical parameter involved here is the adsorption coefficient. It can be considered as the ancestor of LSC. We can include thin‐layer chromatography, which differs from LSC in the distribution of the stationary phase but which proceeds from the same separation principle.

      Liquid/liquid chromatography (LLC)

      The stationary phase and the mobile phase are both liquids. The stationary phase is most often a polymer fixed on an inert support, generally by grafting active species with covalent bonds, ensuring a long lifespan for these columns. The separation relies upon the partition coefficient of the solutes between the two phases (a phenomenon comparable to a liquid–liquid extraction in a separatory funnel, a well‐known process in laboratories). This category also includes centrifugal partition chromatography, which is used in preparative chromatography.

      Ion chromatography (IC)

      Even though the stationary phase is considered to be a solid, the separation principle is different from LSC. Reserved for ionic species, this technique is not based on the adsorption phenomenon, as in LSC, but rather on exchanges between the ions to be separated and those in the stationary phase (anion or cation exchange


Скачать книгу